Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Fluke 863: CAL ADJ RS-232C /5520 INSTRUMENT: Fluke 865: CAL ADJ RS-232C /5520 INSTRUMENT: Fluke 867: CAL ADJ RS-232C /5520 DATE: 21-Aug-98 AUTHOR: Fluke Corporation REVISION: $Revision: 1.2 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 1 NUMBER OF LINES: 339 CONFIGURATION: Fluke 5520A ============================================================================= # # Source: # Fluke 86X Calibration Procedure, Preliminary, February 1995 # # Compatibility: # 5500/CAL 5.0 or MET/CAL 5.0 or later # # Subprocedures: # None # # Required Files: # ini_val.exe Executable: Gets default port for UUT from METCAL.INI # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5520A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # # The 90 day specifications of the 5520A are used in TUR computations. # # The following describes the use of the [I] special constructions # used in the PORT FSC in this procedure: # # SF (Set Function) command # # 6.003 PORT SF 9[13][I][D1000] # # [I] reads the acknowledge response # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- W 1.002 HEAD EQUIPMENT SETUP 1.003 DISP [32] WARNING 1.003 DISP HIGH VOLTAGE is used or exposed during the performance 1.003 DISP of this calibration. DEATH ON CONTACT may result if 1.003 DISP personnel fail to observe safety precautions. 1.004 DISP Plug external power adapter into UUT to insure stable 1.004 DISP power while calibration is being performed. 1.005 DISP If at any time, you wish to abort the calibration, 1.005 DISP select the MET/CAL Terminate option and "Abort CAL" 1.005 DISP (Soft Key 5) on the UUT. 1.006 DISP 10 minute warm-up time requirement. 1.006 DISP Ambient temperature: 18C - 28C. 1.006 DISP Relative humidity: less than 90%. 1.007 DOS ini_val metcal.ini startup port 1.008 DISP Connect the UUT to [MEM2]. 1.009 DISP Rotate the UUT function switch to SET UP. 1.009 DISP 1.009 DISP Verify that Baud Rate = 4800 and Parity = None. 1.009 DISP 1.009 DISP If not, use the Next Item (Soft Key 1) and left and 1.009 DISP right arrow keys (Soft Keys 2 & 3) to select 4800 baud 1.009 DISP and no parity. 1.010 PORT [P4800,N,8,1,X] 1.011 HEAD STEP 1: mVDC CALIBRATION ADJUSTMENT 1.012 DISP Connect the 5520A to the UUT as follows: 1.012 DISP [32] 5520A NORMAL HI to UUT Volt Ohm Diode Input 1.012 DISP [32] 5520A NORMAL LO to UUT COM Input 1.013 DISP Rotate the UUT function switch to mVDC. 1.013 DISP 1.013 DISP Place the UUT in calibration mode by pressing the. 1.013 DISP recessed calibration button located below the power 1.013 DISP adapter plug on the right side of the UUT. 1.013 DISP 1.013 DISP Select Advance when the calibration display appears. 1.014 5520 900mV S 2W # Select Proceed (Soft Key 1). 1.015 PORT [D3000]SF 8[13][I] 1.016 DISP Select Advance when Step 2 appears on the left side of 1.016 DISP the UUT display. 1.017 HEAD STEP 2: mVDC CALIBRATION ADJUSTMENT 1.018 5520 2.9V S 2W 1.019 PORT [D3000]SF 8[13][I] 1.020 DISP Select Advance when Step 3 appears on the left side of 1.020 DISP the UUT display. 1.021 HEAD STEP 3: mVDC CALIBRATION ADJUSTMENT 1.022 5520 290mV S 2W 1.023 PORT [D3000]SF 8[13][I] 1.024 DISP Select Advance when Step 4 appears on the left side of 1.024 DISP the UUT display. 1.025 5520 * S 1.026 HEAD STEP 4: VDC CALIBRATION ADJUSTMENT 1.027 DISP Rotate the UUT function switch to VDC. 1.027 DISP 1.027 DISP Select Advance when the calibration display appears. 1.028 5520 1000V S 2W 1.029 PORT [D3000]SF 8[13][I] 1.030 DISP Select Advance when Step 5 appears on the left side of 1.030 DISP the UUT display. 1.031 HEAD STEP 5: VDC CALIBRATION ADJUSTMENT 1.032 5520 290V S 2W 1.033 PORT [D3000]SF 8[13][I] 1.034 DISP Select Advance when Step 6 appears on the left side of 1.034 DISP the UUT display. 1.035 HEAD STEP 6: VDC CALIBRATION ADJUSTMENT 1.036 5520 29V S 2W 1.037 PORT [D3000]SF 8[13][I] 1.038 DISP Select Advance when Step 7 appears on the left side of 1.038 DISP the UUT display. 1.039 5520 * S 1.040 HEAD STEP 7: mADC CALIBRATION ADJUSTMENT 1.041 DISP Connect the 5520A to the UUT as follows: 1.041 DISP [32] 5520A AUX HI to UUT mA/uA Input 1.041 DISP [32] 5520A AUX LO to UUT COM Input 1.042 DISP Rotate the UUT function switch to mA/uA. 1.042 DISP 1.042 DISP Select Advance when the calibration display appears. 1.043 5520 290mA S 2W 1.044 PORT [D3000]SF 8[13][I] 1.045 DISP Select Advance when Step 8 appears on the left side of 1.045 DISP the UUT display. 1.046 HEAD STEP 8: mADC CALIBRATION ADJUSTMENT 1.047 5520 29mA S 2W 1.048 PORT [D3000]SF 8[13][I] 1.049 DISP Select Advance when Step 9 appears on the left side of 1.049 DISP the UUT display. 1.050 HEAD STEP 9: mADC CALIBRATION ADJUSTMENT 1.051 5520 2.9mA S 2W 1.052 PORT [D3000]SF 8[13][I] 1.053 DISP Select Advance when Step 10 appears on the left side of 1.053 DISP the UUT display. 1.054 HEAD STEP 10: mADC CALIBRATION ADJUSTMENT 1.055 5520 290uA S 2W 1.056 PORT [D3000]SF 8[13][I] 1.057 DISP Select Advance when Step 11 appears on the left side of 1.057 DISP the UUT display. 1.058 HEAD STEP 11: mADC CALIBRATION ADJUSTMENT 1.059 5520 29uA S 2W 1.060 PORT [D3000]SF 8[13][I] 1.061 DISP Select Advance when Step 12 appears on the left side of 1.061 DISP the UUT display. 1.062 5520 * S 1.063 HEAD STEP 12: ADC CALIBRATION ADJUSTMENT 1.064 DISP Connect the 5520A to the UUT as follows: 1.064 DISP [32] 5520A 20A to A Input 1.064 DISP [32] 5520A AUX LO to UUT COM Input 1.065 DISP Rotate the UUT function switch to A. 1.065 DISP 1.065 DISP Select Advance when the calibration display appears. 1.066 5520 10A HC S 2W 1.067 PORT [D3000]SF 8[13][I] 1.068 DISP Select Advance when Step 13 appears on the left side of 1.068 DISP the UUT display. 1.069 5520 * S 1.070 HEAD STEP 13: ADC CALIBRATION ADJUSTMENT 1.071 DISP Connect the 5520A to the UUT as follows: 1.071 DISP [32] 5520A AUX HI to A Input 1.071 DISP [32] 5520A AUX LO to UUT COM Input 1.072 5520 2.9A S 2W 1.073 PORT [D3000]SF 8[13][I] 1.074 DISP Select Advance when Step 14 appears on the left side of 1.074 DISP the UUT display. 1.075 HEAD STEP 14: ADC CALIBRATION ADJUSTMENT 1.076 5520 0.29A S 2W 1.077 PORT [D3000]SF 8[13][I] 1.078 DISP Select Advance when Step 15 appears on the left side of 1.078 DISP the UUT display. 1.079 5520 * S 1.080 HEAD STEP 15: Ohms/nS CALIBRATION ADJUSTMENT 1.081 DISP Connect the 5520A to the UUT as follows: 1.081 DISP [32] 5520A NORMAL HI to UUT Volt Ohm Diode Input 1.081 DISP [32] 5520A NORMAL LO to UUT COM Input 1.082 DISP Rotate the UUT function switch to Ohms. 1.082 DISP 1.082 DISP Select Advance when the calibration display appears. 1.083 5520 2.9MZ S 2W 1.084 PORT [D3000]SF 8[13][I] 1.085 DISP Select Advance when Step 16 appears on the left side of 1.085 DISP the UUT display. 1.086 HEAD STEP 16: Ohms/nS CALIBRATION ADJUSTMENT 1.087 5520 290kZ S 2W 1.088 PORT [D3000]SF 8[13][I] 1.089 DISP Select Advance when Step 17 appears on the left side of 1.089 DISP the UUT display. 1.090 5520 * S 1.091 HEAD STEP 17: Ohms/nS CALIBRATION ADJUSTMENT 1.092 DISP Connect the 5520A to the UUT as follows: 1.092 DISP [32] 5520A SENSE HI to UUT Volt Ohm Diode Input 1.092 DISP [32] 5520A SENSE LO to UUT COM Input 1.092 DISP [32] 5520A NORMAL HI to UUT Volt Ohm Diode Input 1.092 DISP [32] 5520A NORMAL LO to UUT COM Input 1.093 5520 29kZ S CW 1.094 PORT [D3000]SF 8[13][I] 1.095 DISP Select Advance when Step 18 appears on the left side of 1.095 DISP the UUT display. 1.096 HEAD STEP 18: Ohms/nS CALIBRATION ADJUSTMENT 1.097 5520 2.9kZ S CW 1.098 PORT [D3000]SF 8[13][I] 1.099 DISP Select Advance when Step 19 appears on the left side of 1.099 DISP the UUT display. 1.100 HEAD STEP 19: Ohms/nS CALIBRATION ADJUSTMENT 1.101 5520 290Z S CW 1.102 PORT [D3000]SF 8[13][I] 1.103 DISP Select Advance when Step 20 appears on the left side of 1.103 DISP the UUT display. 1.104 5520 * S 1.105 HEAD STEP 20: Ohms/nS CALIBRATION ADJUSTMENT 1.106 DISP Connect the 5520A to the UUT as follows: 1.106 DISP [32] 5520A NORMAL HI to UUT Volt Ohm Diode Input 1.106 DISP [32] 5520A NORMAL LO to UUT COM Input 1.107 5520 3.5MZ S 2W 1.108 PORT [D3000]SF 8[13][I] 1.109 DISP Select Advance when Step 21 appears on the left side of 1.109 DISP the UUT display. 1.110 HEAD STEP 21: Ohms/nS CALIBRATION ADJUSTMENT 1.111 5520 350kZ S 2W 1.112 PORT [D3000]SF 8[13][I] 1.113 DISP Select Advance when Step 22 appears on the left side of 1.113 DISP the UUT display. 1.114 5520 * S 1.115 HEAD STEP 22: VAC CALIBRATION ADJUSTMENT 1.116 DISP Rotate the UUT function switch to VAC. 1.116 DISP 1.116 DISP Select Advance when the calibration display appears. 1.117 5520 1000V 1kH SI S 2W 1.118 PORT [D5000]SF 8[13][I] 1.119 DISP Select Advance when Step 23 appears on the left side of 1.119 DISP the UUT display. 1.120 HEAD STEP 23: VAC CALIBRATION ADJUSTMENT 1.121 5520 100V 1kH SI S 2W 1.122 PORT [D5000]SF 8[13][I] 1.123 DISP Select Advance when Step 24 appears on the left side of 1.123 DISP the UUT display. 1.124 HEAD STEP 24: VAC CALIBRATION ADJUSTMENT 1.125 5520 290V 1kH SI S 2W 1.126 PORT [D5000]SF 8[13][I] 1.127 DISP Select Advance when Step 25 appears on the left side of 1.127 DISP the UUT display. 1.128 HEAD STEP 25: VAC CALIBRATION ADJUSTMENT 1.129 5520 29V 1kH SI S 2W 1.130 PORT [D5000]SF 8[13][I] 1.131 DISP Select Advance when Step 26 appears on the left side of 1.131 DISP the UUT display. 1.132 HEAD STEP 26: VAC CALIBRATION ADJUSTMENT 1.133 5520 2.9V 1kH SI S 2W 1.134 PORT [D5000]SF 8[13][I] 1.135 DISP Select Advance when Step 27 appears on the left side of 1.135 DISP the UUT display. 1.136 HEAD STEP 27: VAC CALIBRATION ADJUSTMENT 1.137 5520 290mV 1kH SI S 2W 1.138 PORT [D5000]SF 8[13][I] 1.139 DISP Select Advance when Step 28 appears on the left side of 1.139 DISP the UUT display. 1.140 HEAD STEP 28: VAC CALIBRATION ADJUSTMENT 1.141 5520 29mV 1kH SI S 2W 1.142 PORT [D5000]SF 8[13][I] 1.143 DISP Select Advance when Step 29 appears on the left side of 1.143 DISP the UUT display. 1.144 5520 * S 1.145 HEAD STEP 29: mAAC CALIBRATION ADJUSTMENT 1.146 DISP Connect the 5520A to the UUT as follows: 1.146 DISP [32] 5520A AUX HI to UUT mA/uA Input 1.146 DISP [32] 5520A AUX LO to UUT COM Input 1.147 DISP Rotate the UUT function switch to mA/uA. 1.147 DISP 1.147 DISP Select Advance when the calibration display appears. 1.148 5520 290mA 1kH SI S 2W 1.149 PORT [D5000]SF 8[13][I] 1.150 DISP Select Advance when Step 30 appears on the left side of 1.150 DISP the UUT display. 1.151 HEAD STEP 30: mAAC CALIBRATION ADJUSTMENT 1.152 5520 29mA 1kH SI S 2W 1.153 PORT [D5000]SF 8[13][I] 1.154 DISP Select Advance when Step 31 appears on the left side of 1.154 DISP the UUT display. 1.155 HEAD STEP 31: mAAC CALIBRATION ADJUSTMENT 1.156 5520 2.9mA 1kH SI S 2W 1.157 PORT [D5000]SF 8[13][I] 1.158 DISP Select Advance when Step 32 appears on the left side of 1.158 DISP the UUT display. 1.159 HEAD STEP 32: mAAC CALIBRATION ADJUSTMENT 1.160 5520 290uA 1kH SI S 2W 1.161 PORT [D5000]SF 8[13][I] 1.162 DISP Select Advance when Step 33 appears on the left side of 1.162 DISP the UUT display. 1.163 HEAD STEP 33: mAAC CALIBRATION ADJUSTMENT 1.164 5520 29uA 1kH SI S 2W 1.165 PORT [D5000]SF 8[13][I] 1.166 DISP Select Advance when Step 34 appears on the left side of 1.166 DISP the UUT display. 1.167 5520 * S 1.168 HEAD STEP 34: AC Amps CALIBRATION ADJUSTMENT 1.169 DISP Connect the 5520A to the UUT as follows: 1.169 DISP [32] 5520A 20A to A Input 1.169 DISP [32] 5520A AUX LO to UUT COM Input 1.170 DISP Rotate the UUT function switch to A. 1.170 DISP 1.170 DISP Select Advance when the calibration display appears. 1.171 5520 10A 1kH SI HC S 2W 1.172 PORT [D5000]SF 8[13][I] 1.173 DISP Select Advance when Step 35 appears on the left side of 1.173 DISP the UUT display. 1.174 5520 * S 1.175 HEAD STEP 35: AC Amps CALIBRATION ADJUSTMENT 1.176 DISP Connect the 5520A to the UUT as follows: 1.176 DISP [32] 5520A AUX HI to A Input 1.176 DISP [32] 5520A AUX LO to UUT COM Input 1.177 5520 1A 1kH SI S 2W 1.178 PORT [D5000]SF 8[13][I] 1.179 DISP Select Advance when Step 36 appears on the left side of 1.179 DISP the UUT display. 1.180 HEAD STEP 36: AC Amps CALIBRATION ADJUSTMENT 1.181 5520 2.9A 1kH SI S 2W 1.182 PORT [D5000]SF 8[13][I] 1.183 DISP Select Advance when Step 37 appears on the left side of 1.183 DISP the UUT display. 1.184 HEAD STEP 37: AC Amps CALIBRATION ADJUSTMENT 1.185 5520 290mA 1kH SI S 2W 1.186 PORT [D5000]SF 8[13][I] 1.187 DISP Select Advance when Step 38 appears on the left side of 1.187 DISP the UUT display. 1.188 5520 * S 1.189 HEAD STEP 38: mVDC CALIBRATION ADJUSTMENT 1.190 DISP Connect the 5520A to the UUT as follows: 1.190 DISP [32] 5520A NORMAL HI to Volt Ohm Diode Input 1.190 DISP [32] 5520A NORMAL LO to UUT COM Input 1.191 DISP Rotate the UUT function switch to mVDC. 1.191 DISP 1.191 DISP Select Advance when the calibration display appears. 1.192 5520 29mV S 2W 1.193 PORT [D3000]SF 8[13][I] 1.194 DISP Select Save Cal. to store new constants or Abort Cal. 1.194 DISP when calibration completes. 1.195 END